Inputs up to 10A and 1000V
0.2% Accuracy
Isolated inputs
Low Power Factor accuracy Harmonic analysis
Total Harmonic Distortion
Low current ranges
USB and IEEE interface
Graphical display
參考價 | 面議 |
更新時間:2018-10-15 11:55:35瀏覽次數:916
聯系我們時請說明是化工儀器網上看到的信息,謝謝!
TRUE RMS/WIDE BANDWIDTH REMOTE CONTROL CURRENT, VOLTAGE, AND POWER OSCILLOSCOPE PHASE, PF, AND VOLT-AMPERES Full Scale Voltage Max. Peak Value Resolution Input Impedance 2.000V 6V 1mV 1.05 MOhms / 30pF 20.00V 60V 10mV 1.05 MOhms /30pF 200.0V 600V 100mV 1.00 MOhms / 30pF 2000 V 3000V 1 V 1.00 MOhms / 30pF The 2000V range is useable to 1000V rms Full Scale Voltage Max. Peak Value Resolution Input Impedance 5.000mA 15mA 1uA 2.0 Ohms 50.00mA 150mA 10uA 2.0 Ohms 500.0mA 1500mA 100uA 0.034 Ohms 5000 mA 15A 1mA 0.034 Ohms EXTERNAL (100mV) 300mV - 100 kOhms 5% - * of Full Scale * - 200% of Full Scale Frequency ?% of Reading+% of Range) % of Reading dc ?0.10 + 0.30) ?.40 5Hz - 250kHz ?0.10 + 0.10) ?.20 250kHz - 500kHz ?0.20 + 0.20) ?.40 *500kHz - 1.0MHz ?0.40 + 0.40) ?.80 POWER AND POWER x 10: Ranges (for any Power Factor) (V-A is the Volt-Ampere product) VOLT-AMPERES: POWER FACTOR: PHASE ANGLE: HARMONICS: THD: FREQUENCY: DISPLAY: DIGITAL INTERFACE: DISPLAY UPDATE: SETTLING TIME: TEMPERATURE RANGE: RELATIVE HUMIDITY WARM-UP TIME: POWER REQUIREMENTS: PHYSICAL: 5% - * of Full Scale V-A * - 200% of Full Scale V-A Frequency ?% of V-A+% of Range) % of V-A dc ?0.20 + 0.30) ?.50 5Hz - 250kHz ?0.20 + 0.10) ?.30 250kHz - 500kHz ?0.40 + 0.20) ?.60 *500kHz - 1.0MHz ?0.80 + 0.40) ?.20
The Model 2335A Wideband Power Analyzer is a precision, high accuracy, high frequency auto-ranging instrument which measures and displays true rms Voltage, true rms Current, and true mean Power over a frequency range from dc to more than 1MHz. Full scale Current and Voltage inputs are typically measured within ?.1% of the amplitude reading to at least 500kHz. The corresponding Power is typically measured to within ?.1% of the input Volt-Amperes to 250kHz and to within ?.2% of the input Volt-Amperes to 500kHz for loads having any Power Factor. MULTI-FUNCTION
In addition to the primary functions, the Model 2335A measures the input Frequency, the phase angle between the current and voltage inputs, and the harmonic amplitudes and phases relative to the fundamental phase, up to 50, for both the current and voltage inputs. It also calculates the Power Factor, the Volt-Amperes, and the Energy (if it has been selected). These functions may be displayed or read over the IEEE-488 or USB interface.
Five digits or resolution combined with excellent phase matching between the current and voltage channels make the Model 2335A an exceptionally good instrument for making low power factor measurements up to 1MHz. This ability makes the instrument ideal for high frequency core loss measurements which have inherently low power factors.
The Model 2335A allows broadband and high accuracy measurements of both sinusoidal and highly distorted wave shapes. The Current, Voltage, Power, and Power Factor uncertainties to 1MHz are far better than any other Watt Meter or Power Analyzer, or for that matter, with respect to Current or Voltage, almost all conventional multimeters.
The Model 2335A has full scale Power ranges from 1.0000mW to 10000W. With external shunts or current to voltage transducers, the upper range may be extended by a factor of ten or one hundred. Full scale Voltage from 2.000V to 2000V (usable to 1000V) and full scale Current ranges from 5.000mA to 5.000A (all rms values) cover a wide range of load impedances. Full scale Current and Voltage inputs may have crest factors up to three while smaller inputs may have even higher crest factors. Sinusoidal inputs with rms values of twice the nominal Full Scale value may be measured with no loss in accuracy.POSSIBLE MEASUREMENT USES
The Voltage and Current inputs of the Model 2335A are simultaneously sampled (with 16 bit resolution), converted to digital form, and transmitted via isolated links to the control board on the main chassis. This allows both the Current and Voltage inputs to be compley isolated from each other and from the main chassis ground. The asynchronous sampling frequency is controlled by the system microprocessor in such a fashion that neither it nor any of its harmonics can come close to the measured input frequency or any of its harmonics. This precaution permits sampling at frequencies that are significantly below the Nyquist rate.
The Model 2335A is equipped with an IEEE-488.2 interface and a USB interface which both incorporate all of the IEEE-488.2 Common Commands and Queries as well as the commands and queries required for device operation. Any function that can be entered via the front panel can be controlled via either interface. A menu item selects either the IEEE-488 or the USB interface to be active.
TYPICAL DISPLAYS
SPECIFICATIONS
Accuracy (rms) Current and Voltage (Double the specification for peak readings)
19" Rack Mountable, 3-1/2" high (2U)
Weight: 5kgm (11pounds)
Size: 48.3cm x 8.90cm x 33.0cm (19" x 3.5" x 13")
100 to 260V rms, 47 to 63Hz, 30VA max.
Thirty minutes for all specifications
Operating: 0?/font> C to 40?/font> C
Within Specifications: 23篊 ?5篊
Storage: -40?/font> C to 75?/font> C:
Less than 90%
Fast (160ms), Medium (2s), Slow (4s).
80ms
USB, IEEE-488.2
High resolution graphical TFT
The Frequency of the Voltage or the Current from 5.00Hz to 3275kHz with four digits of resolution and an accuracy of ?00ppm ? digit.
The Total Harmonic Distortion accuracy is ?1dB.
The Harmonic amplitude accuracy is the same as that of the current or the voltage at the frequency of the harmonic, and the resolution is the same as that of the fundamental (the first harmonic).. The accuracy of the harmonic phase angle is three times as graat as the accuracy of the phase angle between the current and the voltage
The resolution of the phase angle between the fundamental current and the fundamental voltage is 0.01? The accuracy is ?.17?up to 250kHz, ?.34癴rom 250kHz to 500kHz, and ?.04?from 500kHz to 1MHz.
Power Factor is calculated as the ratio of Power to Volt-Amperes. It has a range from 0 to ?.0000 and a resolution of .0001 for Volt-Ampere products greater than 7.5% of Full Scale. The resolution decreases as the Volt-Ampere product decreases. The accuracy is ?.002 ?0 digits up to 250kHz, ?.004 ?0 digits from 250kHz to500kHz, and ?.008 ?0 digits from 500kHz to 1000kHz (with option -01. Eight decade Full Scale ranges from 1.0000mW to 10,000W. The ranges are all combinations of a Full Scale Current range multiplied by a Full Scale Voltage range plus a corresponding set of combinations with ten times the sensitivity which occur when the Px10 range is activated. Any Px10 range may be selected when both the Peak Current and the Peak Voltage are less than 0.316 of their Peak Range values. Any PX10 range may be selected when both the Peak Current and the Peak Voltage are less than 0.316 of their Peak Range values.
Resolution 1 part in 10000 of the Full Scale rangeIsolation: A ?00Vpeak may be placed between the LO Voltage Terminal and Chassis.
*Specify option -01
Volt-Amperes are calculated as the product of the rms Current and the rms Voltage. They have the same Full Scale ranges, resolution and accuracy as POWER.
CURRENT AND VOLTAGE:
Measurement of Ultrasonic Equipment of all types and power levels, Ferrite Core Loss, Transformer Core Material, Switching Power Supplies, Fluorescent Lamp Ballasts of all types, Mercury Arc Lamp Circuits, Sodium Lamp Ballasts, Speed Controlled Motors of all types, Efficiency of any device with an electrical input and an electrical output, SCR Controlled Devices of all types, High Frequency and/or Distorted Currents from any source, Voltage Response of any device from DC to 1MHz, and the Characteristics of Electric Auto Drives.